TM 11-6625-2975-40
2-3. GENERAL. (CONT)
Each test is complete and maybe performed Individually, Therefore, you may choose the appropriate
test to verify gross equipment failure or performance degradation of a specific module. However, this
maintenance approach is not recommended. It is best to perform all the tests In sequence. This
systematic maintenance approach wiII ensure that all faults are found and corrected.
WARNING
HIGH VOLTAGE is used during the following tests. DEATH ON CONTACT may result If
personnel fall to observe safety precautions,
Dangerous potentials exist at several points throughout this equipment, when the
equipment Is operated with the case removed. Do not touch exposed connections or
components.
NOTE
In order to obtain proper results, all tests should be performed under the following
environmental conditions: temperature, + 50 to + 104F ( + 10 to + 40C); maximum
relative humidity, 95 percent.
Before starting the performance test sequence, do the Initial adjustments, self-test, and operational
check as described In TM 11-6625-2975-12, If test set falls any portion of these tests, turn to the
troubleshooting section of this manual and follow the steps In troubleshooting chart 2-1.
NOTE
This test checks output and modulation tone frequencies.
TEST EQUIPMENT REQUIRED: Adapter, N plug-to-BNC jack
Frequency Counter HP 5345A
RF Cable RG-58/U
EQUIPMENT SETUP Connect test set to 115 vac power source.
Connect frequency counter to test set as Indicated In the following diagram.
|