TB 43-0001-61-15-1
SECTION II. ACTIVE (OPEN) EIR CASES
3.2 EIRs Still Under Investigation. The following cases are under investigation: N/A
DAWS Calibration of Scales. This item is pending clarification on the data.
SECTION III. INACTIVE (CLOSED) EIR CASES
EIRs Requiring Answers to the Originator Only. None.
Closed EIR Cases.
a. Fixtures with Dial Indicators. Please continue to carefully examine drawings to
determine certification requirements for fixtures. Point of contact is Mr. Greg Rigney, e-mail:
gregory.m.rigney.civ@mail.mil, DSN: 746-7715, commercial: 256-876-7715.
b. Piston Gage System COMPASS Profiles. An e-mail was sent to all USATA
elements with a piston gage system requesting they verify their COMPASS profiles match
the specifications determined by Engineering. Point of contact is Mr. Wade Weaver, e-mail:
wade.h.weaver.civ@mail.mil, DSN: 746-9504, commercial: 256-876-9504.
c. Torque Cells (Model 2351-102). No further information is available. If any issues
develop please contact Mr. Greg Rigney, e-mail: gregory.m.rigney.civ@mail.mil., DSN: 746-
7715, commercial: 256-876-7715.
MINOR ALTERATIONS/INFORMATION
All minor alterations are optional for application to the item indicated at the field
maintenance level specified. Minor alterations are to be undertaken only if adequate skills,
tools, and parts are available. Application of minor alterations will be recorded in the
appropriate equipment record as a routine maintenance action.
4.2 TEGAM 1830A RF Power Meter. SD cards will no longer be supported in the 1830A.
Starting December, 2014, (with firmware version 1.5.0a), user files will be stored in non-
removable memory that has existed in the 1830A design since the product was first
introduced. No new 1830A will have an SD card installed. All 1830As will still be able to be
updated using the LXI interface, and the functionality of the software version 1.5.0a is
completely unchanged from version 1.1.38a. New-Production 1830A are fully compatible
with older versions. Software version 1.5.0a has the same front-panel and network interface
as 1.1.38a. Any automation that checks for version may need to be updated to recognize
version 1.5.0a. TEGAM will release all future software updates with installation packages
that will move storage from SD card to on-board memory, and discontinue use of the SD card.
Users will be directed to remove the SD card upon completion of the update. TEGAM will