Table 5-1. Direct Support Troubleshooting - Continued
Section III. MAINTENANCE OF MAIN HOUSING ASSEMBLY

Test Set, Electrical (0-33Kw Load Bank) Model 1057 Page Navigation
  15    16    17    18    19  20  21    22    23    24    25  
img
. . .
TM5-6625-2694-13&P
Table 5-1.
Direct Support Troubleshooting (cont'd.)
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
Step 2.
Load Selector Switch(s) defective.
Disconnect power from Test Set.
a.
Remove  and  tag  leads  from  switch  terminals.
b.
Attach  ohmmeter leads across  switch terminals top to
bottom.
c.  With  Switch  in on (up) position, ohmmeter should
read 0  Ohms.  With  Switch  in  off (down)  position,
ohmmeter should read infinite.
Refer to
defective
d.
5-21 and
replace
paragraph
Switch(s).
Step 3.
Check Load Resistors.
Disconnect and tag lead from one end of Resistor.
a.
b.
Attach ohmmeter leads to Resistor terminals.
c.  Ohmmeter should read  15  Ohms
for 1 KW Resistors
and 30 Ohms for 500 W Resistors.
defective
d.
and
Refer
to
replace
paragraph 5-8
Resistors.
5.
VARIABLE LOAD CONTROL HAS NO EFFECT ON LOADING.
Measure AC voltage on TB1-1, TB2-1 and TB3-1 with respect to
Step 1.
AC Voltage should vary in direct
LO while varying  the load control.
proportion to load control.
remove
If  any  phase  exhibits  no  variation
and
replace T4.
Refer to paragraph 5-17