TM 11-6625-3056-12
4-6.
MONTHLY PREVENTIVE MAINTENANCE (cont)
Item
Item To Be
No.
Inspected
Procedure
9.
Sync Time
Set controls as follows:
PATTERN
Keep at SHORT
OPERATION
2W-R
CORRELATOR LOOP
Keep at OPEN
DATA ON/OFF
OFF
RATE SELECT
Keep at O (38.4 kb/s)
DISPLAY SELECT
Keep at OFF
Keep DATA OUT connected to DATA IN:
Set DISPLAY SELECT to SYNC TIME MSEC and press RESET:
Display must read 0.
Set DATA ON/OFF to ON:
Display must rapidly advance count (1000 counts per second).
Set OPERATION to 4W:
Display must stop advancing count and must display some number (representing length of
time, in milliseconds, from the time DATA ON/OFF control was set to ON and the time
OPERATION control was set to 4W).
Press RESET:
Display must read 0.
10.
Two-Wire
Set controls as follows:
Operation
PATTERN
Keep at LONG
OPERATION
Keep at 4W
CORRELATOR LOOP
Keep at CLOSED
DATA ON/OFF
Keep at ON
RATE SELECT
Keep at 0 (38.4 kb/s)
DISPLAY SELECT
Keep at ERRORS
Keep DATA OUT connected to DATA IN.
Set OPERATION to 2W-T:
ACTIVITY indicator must extinguish.
Set OPERATION to 2W-R:
END
ACTIVITY indicator must remain unlit.
4-7
|