TM 1-6625-735-14
*Two required.
ROLL ATT Synchro Performance Check.
(Refer to Figure 5-12)
(Refer to Figure 5-12)
Equipment Connections
Ratio Transformer Settings
ROLL ATT Synchro
Control Indications
30 6 0.2
.500000
a. Connect B4 test lead to test
2106 0.2
.500000
set ROLL ATT SIm test poiunt
Z, using B1 adapter. Connect
B5 test lead to ROLL ATT
SIM test point X using B2
adapter. Connect ac voltmeter
(+) lead to ROLL ATT SIM
test point Y using B1 adapter.
270 6 0.2
.500000
b. Connect B4 test lead to
90 6 0.2
.500000
ROLL ATT SIM test point X
using B1 adapter. Connect B5
test lead to ROLL ATT SIM
test point Y using B2 adapter.
Connect ac voltmeter (+) lead
to ROLL ATT SIM test point
Z using B1 adapter.
150 6 0.2
.500000
c. Connect B4 test lead to
3306 0.2
.500000
ROLL ATT SIM test point Y
using B1 adapter. Connect B5
test lead to ROLL ATT SIM
test point Z using B2 adapter.
Connect ac voltmeter (+) lead
to ROLL ATT SIM test point
X using B1 adapter.
5-73
|