TM 55-6695-220-13&P
3-14.
METER- INSPECT/REPLACE.
3-14
This task covers:
Inspect and Replace.
INITIAL SETUP
Personnel Required
Test Equipment
35H Calibration Repair Specialist
ANGSN 286
Tools
Equipment Condition
Aircraft Electrician's
Para. 3-13.
Tool Kit JTK-17
Condition Description
Panel Assembly removed
1. INSPECT.
Inspect meter for bent pointer and damaged zero adjust.
2. REPLACE.
a. Removal.
(1) Use 3/8 inch open-ended wrench to remove two hex nuts securing circuit leads to meter.
(2) Use wrench and small screwdriver to remove six hex nuts securing meter to panel (See Figure 3-9).
b.
Installation.
(1) Secure replacement meter with six hex nuts.
(2) Replace two circuit leads, observing that red colored wire belongs to positive terminal.
GO TO NEXT PAGE
3-24
|