TO 33D2-10-10-41
TM 55-6695-217-13
WARNING
c. Accomplish a leakage test of the assembled test
equipment.  The leakage rate of the test equipment
Zinc Chromate primer is toxic to skin,
must be subtracted from the leakage rate experienced
eyes and respiratory tract. Us In a well
after the tester is included in the test setup.
ventilated area.
Avoid prolonged
d. Include tester In test setup by attaching the
breathing of vapors.  Avoid eye and
supply connection line to one inlet of the tester tee inlet
repeated skin contact.  Keep away
fitting. Make sure the other inlet Is securely capped.
from sparks and flames.
e. Open line shut-off valve and then the oxygen
supply shut-off valve.
Paints,
primers,
lacquers
and
varnishes must be handled carefully
f. Adjust the oxygen supply pressure regulator
and used only in a well ventilated
until line pressure gauge indicates 450 psi.
approved  area.
Avoid  prolonged
g. Close line shut-off valve and note the time. At
breathing of vapors.  Avoid eye and
the end of 7 minutes the line pressure gauge shall
repeated skin contact.  Keep away
indicate 400 psi minimum. If the reading is less than
from spark and flames.
400 psi, leakage within the oxygen system of the tester
is indicated. Use Leak Tec, manufactured by American
c. Repair of Finish. Touch up finish as required
Gas & Chemical Co., (MIL-L-25567), NSN 6850-00-621-
with zinc chromate primer, MIL-P-6889, and black semi-
1850, or soap and water solution around each
gloss enamel, MIL-E-7729.
connection to determine the location of leak.
d. Assembly.
Assemble tester in reverse of
disassembly order as determined by inspection of
5-6.  CALIBRATION.  The tester does not require
Illustrated Parts Breakdown illustration and parts list.
calibration; however, some components of the tester
may require calibration. The oxygen cylinder pressure
gauge and oxygen regulator may be calibrated as
5-5. LEAKAGE TEST. Leakage test shall be made at
follows:
atmospheric pressure of 28 to 32 inches of mercury at
a. Oxygen Cylinder Pressure Gauge.  Compare
temperature of 77 degrees F, plus or minus 18 degrees
pressure reading of suspect gauge with those of a
F, and at a relative humidity of 80 percent or less. Tests
known accurate gauge, with both connected to same
performed with atmospheric pressure or temperature
oxygen supply. If the pointer of oxygen pressure gauge
substantially different from these values shall have the
under test is within 10 percent of full scale with pressure
proper allowance made for the change In instrument
applied, and the pressure gauge indicates zero pressure
readings.  Oxygen used for test purposes should be
when the oxygen pressure is removed, the gauge may
commercial grade breathing oxygen or per Federal
be regarded as calibrated.
Specification BB-O-925, Type 1, Grade A. Perform
leakage test as follows:
b. Oxygen Regulator. Type A14 oxygen regulators
may be calibrated by those activities which possess the
a. Accomplish operational checkout for oxygen
required equipment to meet the limits specified in TO.
system to assure tester is functional.
15X6-3-2-23, paragraphs 3.8a, 3.8b (10,000 ft only), and
b. Assemble test setup consisting of an external
4.28b.  If the oxygen regulator fails to meet these
oxygen supply, an oxygen shut-off valve, an oxygen
requirements, it should be replaced by qualified
supply pressure regulator set to deliver 450 psi, an
personnel. The leakage test outlined in paragraph 5-5
oxygen pressure regulator gauge, a line shut-off valve, a
shall be performed on the tester oxygen system after
line pressure gauge calibrated to read in 1 pound
any repair or replacement of parts within that system is
divisions and a suitable line to connect to the tester.
completed.
Change 4 5-2