TM 55-4920-430-13
ing the following AC/DC VOLTMETER tests. Adjustment procedures are in para 4-36.
a. Set TESTER POWER switch S1 to ON.
switch is at AC. Check that the METER RANGE switch is at 200V.
NOTE
The following test will be performed using a variable ac power supply,
50mv to 100 volts.
METER HI jack J8. (See fig. 4-6.) Connect the other terminal of the ac power supply and the negative
test lead of the multimeter to EXT METER LO jack J9. Set the digital multimeter to read ac volts.
shall read as listed below.
AC/DC
INPUT
METER
VOLTMETER READING (AC)
SIGNAL LEVEL
RANGE SWITCH
200 MV
49.1 to 50.9 mV
50.0 mV
98.6 to 101.4 mV
100.0 mV
148.1 to 151.9 mV
150.0 mV
187.7 to 192.3 mV
190.0 mV
flashing
220.0 mV
2V
0.491 to 0.509 V
0.500 V
0.986 to 1.014 V
1.000 V
1.481 to 1.519 V
1.500 V
1.877 to 1.923 V
1.900 V
flashing
2.200 V
20 V
4.91 to 5.09 V
5.00 V
9.86 to 10.14V
10.00 V
14.81 to 15.19V
15.00 V
18.77 to 19.23 V
19.00 V
flashing
22.00 V
200 V
49.1 to 50.9 V
50.0 V
98.6 to 101.4 V
100.0 V
|