(4) Control transformer.
1-6. Tabulated Data
a. Identification. The test set has 10 major identifica-
Manufacturer . . . . . . . . . . . Essex Electro Engineers, Inc.
Dual Voltage . . . . . . . . . . . 120/240 primary;
tion plates. The information contained on these plates
26 secondary
is listed below.
Rating . . . . . . . . . . . . . . . 120 VAC input; 26 VAC out-
(1) Identification plate. The identification plate is
put; 240 VAC input; 26
attached to the left of the control panel door. It lists
VAC output
Capacity . . . . . . . . . . . . . 100 volt amps
serial number, contract number, stock number, model
Phase . . . . . . . . . . . . . . . . Single
number, shipping weight and other pertinent unit
(5) Variable transformer.
data.
Manufacturer . . . . . . . . . . . Essex Electro Engineers, Inc.
(2) Ground plate. The plate is located on the load
Type . . . . . . . . . . . . . . . . Stepless output, manually
terminal post directly below the grounding bolt (fig.
adjusted three gang 240/
120 V input, 0-240 output
(3) High voltage connector plate. This plate is lo-
Rating . . . . . . . . . . . . . . . 0 to 3 KW, 3-gang, 3-phase;
manually controlled 240/
cated on the load terminal post directly above the
120 V
240/416 volt receptacle (fig. 2-2).
(6) Contactor.
(4) Low voltage connector plate. This plate is lo-
Manufacturer . . . . . . . . . . . Essex Electro Engineers, Inc.
cated on the load terminal post directly below the
Rating . . . . . . . . . . . . . . . 3 pst, 200 amp, 240 V, AC,
120/208 volt receptacle (fig. 2-2).
24 VDC coil
(5) Instruction plate. This plate is attached to the
Type . . . . . . . . . . . . . . . . 3 pst (Pole, single throw) 24
inside of the left side door. It explains the operating
VDC, 200 A
procedure for the load bank.
(7) Fan motor.
(6) Schematic plate and wiring plate. The electri-
Manufacturer . . . . . . . . . . . Essex Electro Engineers, Inc.
cal schematic plate is located on the inside of the left
Type . . . . . . . . . . . . . . . . 24 VDC
Rating . . . . . . . . . . . . . . . 1/16 HP minimum at 2100
side door. It illustrates the complete unit wiring.
RPM
(7) Load switch plate. The plate identifies the load
Type . . . . . . . . . . . . . . . . 24 VDC
switches it is attached on the control panel (fig. 2-l).
Duty . . . . . . . . . . . . . . . . Continuous
Degree enclosure . . . . . . . . Full
(8) Master switch plate. This plate identifies the
Mounting . . . . . . . . . . . . . Within power absorber
master switch located on the control panel (fig. 2-l).
(8) Rectifier.
(9) Variable load control plate. This plate iden-
Rating . . . . . . . . . . . . . . . 6 AMP DC
tifies the variable load control. An arrow indicates a
Phase . . . . . . . . . . . . . . . . 3
clockwise turn of the control to increase the load. It is
(9) Voltage, sensing module.
located on the control panel (fig. 2-l).
Rating . . . . . . . . . . . . . . 3 pole normally closed, E
(10) Voltage selector plate. This plate identifies the
AMP contact will open with
an applied input 31 to 39
voltage selector switch for a high voltage 240/416 V or a
AC
low voltage 120/208 V condition.
(10) Test set capacity.
I
b. Tabulated Data. The following is a tabulated
Continuous rating . . . . . . . . 120/208 or 240/416 volt, 3
listing of major components of the Model A427
phase, 4 wire, or 120 or
and Model A427B test sets. Although the physical
240 volt, 1 phase, 2 wire
description of all of these components is the same
at 50-1000 cps
Fixed loads . . . . . . . . . . . . 3, 6, 9, 12 or any combina-
for both models, only the variable transformer is
tion up to 30 KW in multi-
physically and electrically identical on both units
pies of 3
that they may be interchanged.
Variable loads . . . . . . . . . . 0 to 3 KW combined fixed
(1) Test set.
and variable control of lost
between 0.5 KW and 33 KV
Manufacturer . . . . . . . . .
.
.
Essex Electro Engineers, Inc.
single phase or balanced
A427 & A427B
Model . . . . . . . . . . . . . .
.
.
phase operation with fixed
Type . . . . . . . . . . . . . .
.
.
Portable skid mounted
and variable load switche
Power . . . . . . . . . . . . . .
.
.
From connected power unit
in proper condition.
(2) Power absorbers.
Temperature limits . . . . . . . 25 F. (-4 C.) to +125 F. (52
Manufacturer . . . . . . . . . . .
Essex Electro Engineers, Inc.
C.) at relative humidity up
Type . . . . . . . . . . . . . . . .
Circular, fan cooled
to 100%
Rating per Absorber . . . . . .
11 Kilowatts
( KW)
Altitude range . . . . . . . . . . Sea level to 8,000 feet (240
Resistive Elements . . , , . . .
4 Elements: 9, 1, 4, KW
meters)
(3) Rectifier.
Acceleration limits . . . . . . . 2.5 g momentarily applied to
any 3 major axis; in both
Manufacturer . . . . . . . . . . . Essex Electro Engineers, Inc.
directions
Type . . . . . . . . . . . . . . . . Selenium (Model A427)
Change 1
|