TM
11-6625-467-34
input connects to the front panel connector RF VM1.
not related to the 30-megahertz oscillator). Note
By the use of an RF jumper cable, the 30 MC jack is
that R2-3 receives an input from S6B-4. Refer to
c o n n e c t e d to the RF amplifier/detector circuit
through the RF VM1 jack.
CUIT SELECTOR D) has an input of +27.5 volts dc.
(3) Next the output of the RF amplifier/detec-
In position 3, the movable contact connects +27.5
tor circuit must be connected to the TEST METER.
volts S6B-4 which is connected to terminal 3 of
The RF amplifier/detector circuit output (XA13-P)
HIGH FREQ AMPL control R2.
connects to three places: S4A-5, S4A-6, and S4A-7.
(2) So far, note that if CIRCUIT SELECTOR D
is in position 3 and CIRCUIT SECLECTOR C is in
Refer to figure FO-3 and note that the output of
position 7 or CIRCUIT SELECTOR A is in position
S5A connects to S3A-2 and the output of S4A con-
nects to S3A-1. The output of switch S3 connects to
O, +27.5 volts is connected to the HIGH FREQ
A M P L control which acts as a variable voltage
the TEST METER through the PRESS TO TEST
divider. The movable arm of the voltage divider con-
switch. From the above it is seen that, with a pro-
nects to the input of the 30-megahertz oscillator
gram setting of 04X3 or 05X3, the 30-megahertz
through one or two paths, S3 or S5. Therefore, with
oscillator is activated and connected to the TEST
METER. Position 6 of S5 cannot be used because S5
a CIRCUIT SELECTOR program setting of 0XX3 or
XX73, the 30-megahertz oscillator operates and pro-
must be in position 7 to connect the HIGH FREQ
AMPL control output to the 30-megahertz oscillator
duces a 30-megahertz output at XA1l pin C. Note
that XA1l pin C connects directly to J25, a front
if S3 is not in position O. Pushing PRESS TO TEST
panel connector labeled 30 MC. The RF amplifier/
switch S2 connects the RF amplifier/detector cir-
detector circuit must be used to measure the ampli-
cuit output to the TEST METER. The 30-megahertz
tude of the 30-megahertz output signal. In figure
oscillator can now be set for any specified reading
by adjusting the HIGH FREQ AMPL control.
2-83
|