TM 11-6625-2976-40
2-14. VORBEARING SELECTOR CONTROL ALINEMENT. (CONT)
PROCEDURE (CONT)
4. Rotate VOR bearing selector control clockwise, stopping at each control setting. Verify
oscilloscope waveform at each control setting matches appropriate square wave in diagram
above.
2-15. ATTEN CONTROL ALINEMENT.
TEST EQUIPMENT REQUIRED
Dual-Volt Power Supply, HP PS503A
Power Meter, HP 432A with H75-478A
EQUIPMENT SETUP
Connect equipment as indicated in the following diagram.
NOTE
Dual-volt power supply to test set J2 pin connections shown in preliminary procedure
(page 24).
2-57
|