TM 11-6625-2975-12
3-9. OPERATIONAL CHECK. (CONT)
INITIAL SETUP
Perform self-test operation (para 2-3, steps 1 through 5),
OPERATIONAL CHECK PROCEDURE
The following table gives the operational check procedures or minimum performance test for test set.
Perform each step in action column. Next, see the indication column for the expected response. The
indication column also refers the operator to the appropriate repair or replacement procedure.
ACTION
INDICATION
Test set display status should be as follows
1. Set rf output level to -50 dB mW and
for each selected frequency:
sequentially select frequencies of
74.60,75.00, and 75.40 MHz.
INDICATION
DISPLAY
RDL/DDM/MOD
95.0 MOD (%)
AUDIO STATUS
STANDARD
FREQUENCY
(frequency selected) MHz
MODE
MB
-50 dB mW
RF LEVEL
PH LOCK, LEVEL CAL
RF STATUS
400 Hz
TONE SELECT
FUNCTION/
No keys lit
CONTROL
DATA ENTRY
No keys lit
RF SELECT
RF FREQ key lit
If display status is not indicated. as shown
above, refer test set to higher level of
maintenance.
1300 Hz key lamp should light. If not, see
2. Press 1300 Hz key.
If not, refer test set to higher level of
maintenance.
3000 Hz key lamp should light. If not, see
3. Press 3000 Hz key.
If not, refer test set to higher level of
maintenance.
No changes should occur on front panel. If
4. Sequentially press 30 HZ VAR, 9960 FM,
9960 Hz, 1020 Hz, 90 Hz, and 150 Hz
any change occurs, refer test set to higher
level of maintenance.
keys.