2-4. OPERATING PROCEDURES. (CONT)
7. Press STEP D RDL D DDM key until RDL/DDM/MOD display indicates next step listed in
NOTE
Each time STEP D RDL D DDM key is pressed, test set switches to a new DDM step value
until .000 is reached. Pressing STEP D RDL D DDM key again will repeat procedure.
Holding STEP D RDL D DDM key depressed does not provide continuous step increases.
8. Repeat step 7 for each STD DDM step listed in table 2-12 until .000 is indicated by
RDL/DDM/MOD display.
EL9LX041
NOTE
Each change of DDM step requires pressing STEP D RDL D DDM key. DDM will step In
U/L (150 Hz) direction unless D/R (90 Hz) indicator is lit. If D/R (90 Hz) indicator is lit,
DDM steps in D/R (90 Hz) direction until maximum deviation is reached, and next step
returns DDM to .000 or beam center. At beam center, DDM returns to U/L (150 Hz)
signal and indication and remains in U/L (150 Hz) condition until D/R (90 Hz) is again
selected.
2-33