TM 11-6625-2937-13
Table 5-2. Control-Interface Unit Troubleshooting--Continued
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
13.
TEST NUMBER INDICATOR DOES NOT INDICATE PROPERLY (SOME SEGMENTS DARK, ETC.).
Check circuit card assembly A1 (para 5-29e).
Replace circuit card assembly Al (para 5-15).
14.
POWER ON INDICATOR LAMPS DSI LIGHTS, MODE/POWER INDICATOR LAMPS DO NOT LIGHT, AND
TEST NUMBER INDICATOR IS COMPLETELY DARK.
Step 1. Check + 5 V output of power supply AIPSI as follows:
(a) Plug STE extender card into connector AR position.
(b) Check for +5.0O0.I V between pin 43 (+) and pin 44 (-) on extender card.
(c) If voltage is correct, continue to step 2.
(d) If voltage is out of adjustment, adjust AIPSI (para 5-10).
(e) If voltage is missing check power supply terminals 3 (+) and 4 (-) for +5.0:0.l V. Replace defective power
supply (para 5-I8a).
Step 2. Check MODE/POWER indicator lamps and TEST NUMBER indicator for defective wiring (fig. FO-2,
and table 5-4). Replace defective wiring.
15.
TEST ADAPTER FAILURE.
Step 1 If you suspect one or more of the test adapters is not functioning properly, perform continuity tests using
the schematics in figure FO-4.
Step 2. Non-repairable items. Replace with good adapter.