TM 11-6625-2884-30/NAVAIR 16-35TS3615-2
TABLE 1-5. BTS POWER SUPPLY INFORMATION (Continued)
FUNCTION AND CHARACTERISTICS
Supplies -15 V to digital and RF circuits in 1A3/1A4.
The characteristics of the power supply are:
INPUT VOLTAGE 105-125 V ac, 50-420 Hz
OUTPUT VOLTAGE -14.85 to -15.15 V
OUTPUT CURRENT 0.1 amps
REGULATED - Yes
OVERVOLTAGE PROTECTION - NO
SHORT CIRCUIT PROTECTION - Yes
FUNCTION
INDICATOR
JAM/TEST/GO/NO-GO - When pressed, initiates built-in-test equip-
1A1S4/D11
ment (BITE) in LRU-1 under test.
JAM - Lights when LRU-1 under test has acquired a simulated threat
and has activated ECM.
TEST - Lights when BITE in LRU-1 under test has been initiated.
GO - Lights when LRU-1 under test is operating correctly.
NO-GO - Lights when a fault is detected in LRU-1 under test.
Lights when the LRU-1 under test detects a fault in the
INOP -
pulsed TWT (AR1) or if a high temperature condition is
sensed.
POWER - Lights when the +15 V power supply is enabled, showing
breaker-POWER 115 V ac is turned on.
SORTER LOAD COMMAND (CHANNEL 1) - Lights when the first pulses are
received in LRU-1 under test via supervisory control
section of the CM set.
SLO LOAD COMMAND (CHANNEL 1) - Lights when a PRI is acquired by
1A2DS2
sorter No. 1 in LRU-1 under test.
THREAT A (CHANNEL 1) - Lights when the A-W or A-F threats are
1A2DS3
detected in channel 1 in LRU-1 under test
CON SCAN (CHANNEL 1) - Lights when a conical scan modulation
pattern generated by the BTS is detected in channel 1 in
the LRU-1 under test.