Verify Test State 1:16_2 on the CADU is selected.
(1)
On the Electronic Test Set position ACC Signal Select to 1:16. Press DO twice on the CADU.
(2)
The procedure will repeat as per step a).
(q)
Verify Test State 1:64_2 on the CADU is selected.
(1)
On the Electronic Test Set position ACC Signal Select to 1:64. Press DO twice on the CADU.
(2)
The procedure will repeat as per step (j).
(r)
Verify Test State TEST on the CADU is selected.
(1)
On the Electronic Test Set position ACC Signal Select to 1:64. Press DO twice on the CADU.
(2)
The procedure will repeat as per step 0).
(s)
Select Finish and press DO.
(t)
Select Main Menu and press DO.
(u)
Press F2.
(v)
Select Summary Displays. Press DO.
(w)
Proceed to CALIBRATED DATA.
(4) CALIBRATED DATA
(a)
Compare the Summary Display for system under test with figure 2-4 (the values in parentheses are the allowable
tolerances). If the DAU exhibits faulty tolerance readings refer to the Troubleshooting section in chapter 4.
NOTE
Recommend using printer to print out summary displays to assist in summary data review.
(b)
Press QUIT twice to return to main menu.
(5) UTD Test
(a)
The UTD test is limited on the Electronic Test Set. Calibration is limited to:
(1)
Ensuring the UTD infrared light illuminates.
(2)
Ensuring the UTD registers measurement.
2-11
|