TM 1-6625-736-13&P
(g)
Select NEW. Press DO. When DO is pressed, an Entry Form will appear in the display allowing entry of a new
tail number. When calibrating the DAU with the Electronic Test Set, enter the alphanumeric serial number of the
CADU using the F1 and F2 function keys. The F1 and F2 keys scroll through the allowed characters set. That
set is the blank character, 0 - 9, A - Z, and a - z. F1 scrolls forward through the list and F2 scrolls backward.
Selection of a character is accomplished by scrolling to characters and moving to a different position using the
LEFT and RIGHT keys. The rightleft arrow keys move the cursor to the next character or previous character in
the serial number. After the desired serial number has been entered, press DO to store the new serial number
and return to the Main Operations Menu. The QUIT key can be employed at any time to cancel the serial
number entry and return to the Main Operations Menu.
(h)
Select CAL. Press DO.
NOTE
The 1: 1 attention range on the Electronic Test Set is not used for 54 mV/g accelerometer testing.
(i)
Verify Test State 1:4_1 on the CADU is selected.
(1)
On the Electronic Test Set position ACC Signal Select to 1:4. Press DO twice on the CADU.
(j)
The CADU will transfer to the MEASURING screen and automatically process the signals to the DAU.
Successful completion of the measurement data results in a screen that reads:
Change 2-9
|