TABLE NO. 3-46. Multimeter Does Not Indicate Less Than 1.0 Ohm With FAULT MONITOR I LIMIT
2 Switch Placed To INHIBIT And Does Not Indicate Greater Than 100K Ohms With FAULT MONITOR
I LIMIT 2 Switch Placed To NORM. (Cont)
TEST OR INSPECTION
CORRECTIVE ACTION
Step 1. If multimeter indication is not correct, replace switch S11.
Step 2. If multimeter indication is correct, go to 3.
3. Place FAULT MONITOR I LIMIT 2 control to NORM. Multimeter shall indicate
greater than 100K ohms. Go to 4.
4. If multimeter indication is not correct, replace switch S11.
3-39/(3-40 Blank)
|