Page Title: TABLE NO. 3-43. Multimeter Does Not Indicate 28 6 0.15 Vdc When RATE GYRO TEST ROLL Switch is Pressed and Held To TEST.
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

TM 1-6625-735-14
TABLE NO. 3-43. Multimeter Does Not Indicate 28 6 0.15 Vdc When RATE GYRO TEST ROLL
Switch is Pressed and Held To TEST.
TEST OR INSPECTION
CORRECTIVE ACTION
1. Connect multimeter between ROLL TEST MONITOR and PWR GND test jacks.
Go to 2.
2. Press and hold RATE GYRO TEST ROLL control to TEST. Multimeter shall
indicate 28 6 0.15 vdc.
Step 1. If multimeter indication is not correct, go to 3.
Step 2. If multimeter indication is correct, go to 4.
3. Check switch S18 for continuity.
Step 1. If continuity exists, replace filter FL2.
Step 2. If continuity does not exist, replace switch S18.
4. Release RATE GYRO TEST ROLL control. Multimeter shall indicate 0.0 vdc.
Go to 5.
5. If multimeter indication is not correct, replace switch S18.
TABLE NO. 3-44. Multimeter Does Not Indicate 28 6 0.15 Vdc When RATE GYRO TEST YAW
Switch is Pressed and Held To TEST.
TEST OR INSPECTION
CORRECTIVE ACTION
1. Connect multimeter between YAW TEST MONITOR and PWR GND test jacks.
Go to 2.
2. Press and hold RATE GYRO TEST YAW control to TEST. Multimeter shall indi-
cate 28 6 0.15 vdc.
Step 1. If multimeter indication is not correct, go to 3.
Step 2. If multimeter indication is correct, go to 4.
3. Check switch S17 for continuity.
Step 1. If continuity exists, replace filter FL2.
Step 2. If continuity does not exist, replace switch S17.