TM 55-6695-220-13&P
Section V. MAINTENANCE PROCEDURES
3-10.
PYROMETER TESTER-- INSPECT/REPAIR/CALIBRATE.
3-10
This task covers:
Inspect, Repair, Replace and Calibrate.
INITIAL SETUP
Personnel Required
Test Equipment
35H Calibration Repair Specialist
ANGSN 286
Tools
Equipment Condition
Calibration Repair Specialist's
Para. 3-7d.
Tool Kit JTK-17
Para. 3-14.
Para. 3-17.
1.
INSPECT AND REPLACE.
Inspection and replacement of voltmeter, boards PC-1 or PC-2 as described in paragraphs 3-14. and 3-17.
requires that tester be recalibrated.
2.
CALIBRATE.
a.
Set up test equipment as described in para. 3-7d.
b. Adjust coarse and fine rheostats to make digital multi-meter read 48.89 millivolts.
GO TO NEXT PAGE
3-16
|