TM 55-4920-430-13
(3), Connect the positive terminal of No. 1 dc power supply to one side of the decade resistor. Con-
nect the other side of the decade resistor to J1-NN. Connect the negative side of the power supply to Jl -
MM. Set the decade resistor to 1000 ohms.
(4) Connect the digital multimeter across the decade resistor, positive lead of the multimeter con-
netted to the positive output of the power supply. Set multimeter to the 2 volt DC range.
(5) Adjust the output of the power supply for +0.100 volt dc on the digital multimeter.
NOTE
This applies approximately 0.1 milliampere of input current to J1-
(6) Disconnect digital multimeter from power supply and decade resistor. Connect the digital
multimeter set to measure 20-volt ac to receptacle pins J1-s (line) and J3-AA. The multi meter shall in-
dicate less than 0.03 volt ac.
(7) Connect No. 2 power supply across receptacle pinsJ3-A (+) and J1-C.
NOTE
When a test includes a requirement that an ac voltage be in-phase or
out-of-phase, oscilloscope channel A must be connected across the
digital multimeter. Oscilloscope channel B must be connected to the ac
source and the oscilloscope externally triggered from the ac source.
(8) Adjust the output of No. 2 power supply to +28 volt dc. Monitor the Feedback No. 1 output on
the digital multimeter. Within 1.6 to 2.4 seconds the multimeter shall indicate 3.0 volts ac out-of-phase.
The indication shall rise to 4.45 to 5.05 volt ac, out-of-phase.
(9) Set the ROLL switch to RESET. The digital multimeter shall indicate less than 0.03 volt ac.
( 10) Disconnect the positive lead of the multi meter from receptacle pin J1-s. Connect it to recep-
tacle pin J1-j.
(11) Set the ROLL switch to OPERATE. Monitor the Feedback No. 2 output on the digital mul-
timeter. Within 1.6 to 2.4 seconds, the multimeter shall indicate 1.5 volts ac out-of-phase. The indica-
tion shall rise to 2.18 to 2.58 volts ac, out-of-phase.
(12) Set the ROLL switch to RESET. The multi meter shall indicate less than 0.03 volt ac.
(13) Disconnect the positive lead of the digital multi meter, from receptacle J1-j. Connect it to
receptacle pin J1-m.
(14) Set the ROLL switch to OPERATE. Monitor the Feedback No. 3 output voltage on the
digital multimeter. Within 1.6 to 2.4 seconds, the multimeter shall indicate 1.5 volt ac, out-of-phase.
The indication shall rise to 2.18 to 2.58 volt ac, out-of-phase.
(15) Set the ROLL switch to RESET. The multimeter shall indicate less than 0.03 volt ac.
(16) Reduce the output of the power supplies to 0.
(17) Disconnect No. 2 power supply.
(18) Disconnect the multi meter.
(19) Disconnect No. 1 power supply and decade resistor.