TM 55-4920-412-13&P
n.
Shift Register Drive, ANA-DIG UPDATE, Internal
Clock, Reset Pulse, Clock Inhibits, Test Freq, PIS
Control Auto Dim, DIG Display Update (Oscillator Board
A6), Circuits.
(1)
Set ON-OFF-SIM ONLY to OFF.
(2)
Connect test equipment as shown in
Figure 4-1, unit tester connected.
(3)
Disconnect Oscillator Board (A6) from unit
Test and reconnect extender board to oscillator board.
(4)
Connect OSCILLOSCOPE between test
loops 30 (HI) and 44 (LO) (Shift Register Drive). Set
Power to ON.
(5)
Set INTER CAL SELECT to position 21.
(6)
Set
ANALOG
SET
switch
STOP/SLOW/FAST
to
STOP
and
THEN
SLOW.
Oscilloscope will display positive-going square wave
pulses 4.5 to 5.5, Volts peak to peak (Vp-p). If not, see
Troubleshooting Table 4-32.
(7)
Set ANALOG SET switch to FAST.
(8)
Oscilloscope will display positive-going
square wave pulses 4.5 to 5.5 Vp-p. If not, see
Troubleshooting Table 4-32.
(9)
Place ANALOG SET to STOP position.
Oscilloscope will indicate ground.
(10)
Place ANALOG SET to FAST. Interface
Monitor will reset after counting 1827. If not, see
Troubleshooting Table 4-32.
(11)
Disconnect from Test loop 30 after setting
power to OFF.
(12)
Connect oscilloscope counter to test loop
38 (Digital display Update and clock inhibits). Set Power
to ON.
(13)
Oscilloscope will display negative-going
digital update pulses, 4.5 to 5.5 Vp-p. If not, see
Troubleshooting Table 4-32. Repeat technique (11) and
(12) above for test loop 28 for negative-going update
pulses.
(14)
Set Power to OFF and repeat technique
(11) and (12) above. Disconnect oscilloscope from test
loop 28 and reconnect to each test loops 31,33,35,29,
one at a time. Oscilloscope will display positive-going
clock inhibit pulses 4.5 to 5.5 Vp-p for each connection.
If not, see Troubleshooting Table 4-32.
(15)
Set
power
to
OFF
and
disconnect
oscilloscope from test loop 29.
(16)
Connect oscilloscope to test loop 4
(Analog-Digital Update Pulses).
(17)
Set Power Switch to ON and place Analog
Set switch to SLOW.
(18)
Oscilloscope will display negative going
update pulses between 4.5 and 5.5 Vp-p. If not, see
Troubleshooting Table 4-32. Repeat technique (15)
through (18) above for test loop 9.
(19)
Set Power Switch to OFF and disconnect
from test loop 9 and connect oscilloscope and frequency
counter to test loop 26 (Internal clock).
Table 4-31. Signal Data Converter Section
Unit Tester
Digital Voltmeter Indications (vdc)
Switch Settings
Pin Connections
Min
Max
CLOCK INHIBIT to ANA
J8 Pin 24
45
5.5
CLOCK INHIBIT to DIG
J8 Pin 25
4.5
5.5
CLOCK INHIBIT to DIG
J9 Pin 36
4.5
5.5
CLOCK INHIBIT to ANA
J9 Pin 37
4.5
55
P/S CONTROL fully CW
and P/S CONT to REMOTE
ILS test point
4.5
5.5
(connect jumper between
J3-26 and J3-27)
4-58 Change 2
|