TM 11-6625-3056-12
2-5.
SELF-TEST (cont)
3.
Short
Set controls as follows:
Pattern
PATTERN
SHORT
OPERATION
Keep at 4W
CORRELATOR LOOP
Keep at OPEN
DATA ON/OFF
OFF
RATE SELECT
Keep at 1 (0.6 kb/s)
DISPLAY SELECT
Keep at ERRORS
Keep DATA OUT connected to DATA IN.
ACTIVITY indicator must not be lit.
Set DATA ON/OFF to ON:
ACTIVITY indicator must light.
Set RATE SELECT to positions 2 thru 7.
ACTIVITY indicator must remain lit at each position of RATE SELECT (except 0).
4.
Short
Set controls as follows:
Pattern
Correlator
PATTERN
Keep at SHORT
Loop
OPERATION
Keep at 4W
CORRELATOR LOOP
Keep at OPEN
DATA ON/OFF
Keep at ON
RATE SELECT
Keep at 7 (32 kb/s)
DISPLAY SELECT
Keep at ERRORS
Keep DATA OUT connected to DATA IN.
Press RESET:
Display must read 0.
Press SINGLE ERROR several times:
Display must advance one count each time SINGLE ERROR is pressed.
Set CORRELATOR LOOP to CLOSED and press RESET:
Display must read 0.
Press SINGLE ERROR several times:
Display must advance one count each time SINGLE ERROR is pressed.