2-3.
SYNCHRONIZATION
CORRELATOR
The TS-3981 must be synchronized to the input data for proper
LOOP
operation. The CORRELATOR LOOP control provides two modes
of operation:
OPEN
Synchronization is automatic and requires no operator attention.
Input data bits are continuously loaded into the receiver PR
Generator and compared to the generator output. Any difference
is an error.
The error count will be incorrect if more than one error occurs
every 15 bits.
CLOSED
Synchronization is manual and is achieved by momentarily
setting the CORRELATOR LOOP to OPEN and then to
CLOSED.
After 15 input data bits are loaded into the receiver PR
Generator, the CLOSED mode allows the generator to operate
independently. The input data bits are compared to the generator
output. Any difference is an error.
Loading the 15-stage PR Generator requires only 15 data bits.
Even at the lowest data rate of 600 bits per second, the loading
operation (CORRELATOR LOOP OPEN) requires less than 0.03
seconds.
The receiver generally requires resynchronization whenever the
PATTERN, DATA ON/OFF, or RATE SELECT controls are
changed. A nonsynchronized condition is indicated by a very
rapidly changing ERRORS display.
2-4.
PREPARATION FOR USE
1.
Ground
On rear panel, press in cap on ground lug to expose internal slot.
Connection
Place heavy-gage ground wire not supplied) in slot and release pressure on cap.
Connect other end to suitable ground. In the shelter this ground can be an equipment
rack. Outside the shelter this ground can be a ground rod.
2.
Power
Connect power cable to a 115-volt power source.
Connection