Table 6-2. Test Set Checkout Procedure - Continued
Test instructions
Normal indication
m.
(1)
Connect positive lead to
(1)
Infinite ohms
test point H, common lead to
GRD. Switch S5 to OPEN
position. Set DVM to RX1OK
scale. Switch S4 to BULK
position.
(2)
Connect common lead to test
(2)
Less than 100 ohms
point H, positive lead to GRD.
Set DVM to RX1 scale.
n.
(1)
Connect positive lead to
(1)
Greater than 300K
test point D, common lead
ohms
to GRD. Switch S5 to SHORT
position. Set DVM to RX1OK
scale.
(2)
Connect common lead to test
(2)
Less than 2K ohms
point D, positive lead to GRD.
Set DVM to RX100 scale.
2. Voltage Measurements
NOTE
The following voltage
measurements will be made
with the test set connected
through a variac (see
Maintenance Allocation Chart,
appendix D) into a 115 VAC
(50, 60 or 400 Hz) power outlet.
a. Adjust the variac for 115 VAC.
115 VAC
Set within the accuracy of the DVM.
b. Connect test set power cord into
Red lamp should light
variac. Turn power switch to ON position.