OPERATING PROCEDURES. (CONT)
2-4.
NOTE
The following step will ready DATA ENTRY keys to accept entry of DDM selections.
Press and hold RDL/DDM key until it lights.
2.
NOTE
The following step requires a selected DDM to be entered into test set using DATA ENTRY
keys. The decimal point must be entered explicitly. For example, if 0.040 is to be
entered, enter .04.
3. Using DATA ENTRY keys, enter a standard DDM step from table 2-12.
Press ENTER key once.
4.
NOTE
If RDL/DDM/MOD display flashes on and off, press and hold CLEAR key until display
stops flashing. Then, repeat step 3 and enter correct DDM step.