TM 11-6625-2975-12
2-4. OPERATING PROCEDURES. (CONT)
RF Output Level Decrease
NOTE
The following step will cause
rf output level to decrease to 116 (-dB mw) in 10 db steps.
1. Press and hold FAST DECR key until 116 appears in RF LEVEL display as shown in view A above.
NOTE
The following step will cause rf output level to decrease to 120 (-dB mw) in 1 db steps.
2. Press and hold SLOW DECR key until 120 appears in RF LEVEL display as shown in view B above.
RF Output Reference Level Change
NOTE
The rf output can be read in decibels (-dB mw) or voltage (KV or mv) depending on test
requirements. The following step will cause RF LEVEL display to read in voltage.
2-14
|