TM 11-6625-2975-12
1-13. TEST SET OPERATION. (CONT)
SIGNAL
OUTPUT PARAMETER
CHARACTERISTIC
VOR rf carrier frequency
Selected standard VOR ground station frequency.
30 Hz variable AM/9960 Hz FM (30 Hz reference)
Rf carrier modulation
composite audio signal.
30.0 percent.
Amplitude (%) modulation
000.00 degree.
VOR radial
TO or FROM status indicator
FROM.
-50 db mW. (If the output level is changed after
Output level
initial power on or after self-test, the level
remains at the selected level.)
The rf carrier frequency is variable to greater than 0.025 percent of selected frequency to check
receiver selectivity. The carrier frequency can be varied in either direction at a fast or slow rate
depending on test requirements.
The rf output level is adjustable from -6 to -120 db mW to allow operator selection of output level
required to meet test requirements for receiver response. The output level can be increased or
decreased in either 1 or 10 db steps as required by the operator.
Amplitude (%) modulation of simulated VOR signal is adjustable from 10.0 to 35.0 percent in
O.1-percent increments to check receiver response.
TO/FROM indication is determined by phase relationship of composite audio VOR signal and is always
a FROM signal on initial selection of VOR mode. Through keyboard entry, the FROM signal can be
changed to a TO signal, which phase shifts VOR bearing signal by 160 degrees.
A 1020 Hz audio tone to test the receiver audio circuits can be added to VOR rf output signal at 30-
percent modulation by keyboard entry. Individual control of identification signal allows operator to
add or remove the signal as required during receiver checks.
Either or both of the 30 Hz variable or 9960 Hz FM signals can be removed as required to perform
receiver flag checks.