TEST CONDITIONS
3-11. Unless otherwise specified, all tests will be performed under the fol-
lowing conditions.
Temperature: Room ambient (+25 5C)
Altitude: Normal ground
Humidity: Room ambient (up to 90 percent relative humidity)
Input Power: 105 - 125V, 60 Hz
Warm Up: 5 minutes, associated test equipment 30 minutes
3-12.
The following definitions apply throughout this book:
a. Standard Logic Levels (SLL).
Logical 1 = +2.4 to +5.0 volts
Logical O = O to +O.6 volts
Pulse Width.
Pulse width is time measured between the 50% points of the
b.
waveshape.
c. Pulse Repetition Interval (PRI). PRI is the time between consecutive
pulses, usually given microseconds. PRI is measured from a point
on one pulse to the corresponding point on the next pulse.
d.
O to 3 ohms.
TEST EQUIPMENT USE
3-13.
a. Oscilloscope. Either one of the dual channel vertical inputs shall be
used. Triggering shall be as specified in the test instructions.
Normal inputs/outputs implied unless otherwise
b. Other Test Equipment.
specified.
3-9
|