TM 11-6625-2884-30/NAVAIR 16-35TS3615-2
CONTROL AND POWER DISTRIBUTION (Continued)
b. YIG FILTER Control . The BTS contains a microwave bandpass filter of the
Yttrium Iron Garnet (YIG) type.
(1) 115 VAC,
60 Hz, 10 is applied through input power jack J5 to line
filters
FL1 and FL2 which eliminate conductive EMI from the BTS.
Circuit
Breaker CB2, when enabled, supplies input ac power to
Circuit
Breaker CB3.
(2) Circuit Breaker CB3 is a push-to-reset device located on the front
panel of 1A1. When CB3 is enabled ac power is applied to
YIG Filter Power Supplies PS1, PS2 and PS3.
( 3) Power Supply PS1, when enabled, supplies -15 Vdc 1% with an
output current of 0.1 amps to the YIG filter in the status panel
assembly. Power Supply PS2, when enabled, supplies +15 Vdc 1%
with an output current of 1.0 amps to the YIG filter in the status
panel.
(4) POWER Indicator DS3 lights when +15V Power Supply PS2 is enabled.
(5) Power Supply PS3, when enabled, supplies +10 Vdc 1% with an
output current of 0.12 amps to the YIG FILTER CONTROL switch S3.
S3 is a three decade thumbwheel switch used to tune the YIG filter
to the F1, F2 and F3 frequencies.
(6) The +28 Vdc is applied through INPUT POWER jack J6 to line filter
FL3 and FL4 which eliminate conductive EMI from the BTS. Circuit
Breaker CB1, when enabled, supplies heater power to YIG filter
1A2FL1.
(7) YIG FILTER IN/OUT Switch S2 is used to switch the YIG filter in or
out of the RF Path in the status panel.
OUT position - YIG filter bypassed (no insertion loss)
IN position = YIG filter in circuit (used to measure depth of modu-
lation)
(8) Al 1 signals are routed through P4 to status panel connector 1A2J10.