TM 1-6625-735-14
Turned Fully Counterclockwise And Does Not Indicate 13.5 6 0.15 Vdc When ACTR POSN Control Is
Turned Fully Clockwise. (Cont)
TEST OR INSPECTION
CORRECTIVE ACTION
4. Connect multimeter between ACTR POSN SIM NO. 2 and SIG GND test jacks.
Go to 5.
5. Turn ACTR POSN control throughout its range from fully counterclockwise (ACTR
POSN DN) to fully clockwise (ACTR POSN UP). Multimeter shall indicate -15.0 and
15.0 vdc.
Step 1. If multimeter indication is not correct, go to 6.
Step 2. If multimeter indication is correct, replace switch S1.
6. Connect multimeter between terminal 2 of potentiometer R1 and signal ground.
Go to 7.
7. Turn ACTR POSN control throughout its range from fully counterclockwise (ACTR
POSN DN) to fully clockwise (ACTR POSN UP). Multimeter shall indicate 15.0 and
-15.0 vdc.
Step 1. If multimeter indication is correct, go to 8.
Step 2. If multimeter indication is not correct, replace potentiometer R1.
8. Check switch S1 for continuity.
Step 1. If continuity exists, replace circuit board assembly A1.
Step 2. If continuity does not exist, replace switch S1.
To SIM STAB 1, SIM STAB 2, Or SIM STAB 1&2.
TEST OR INSPECTION
CORRECTIVE ACTION
Replace switch S1.
3-19
|