TM 1-6625-735-14
Is Turned Fully Counterclockwise And Does Not Indicate -7.16 6 0.74 Vdc When CLTV STICK POSN
Control Is Turned Fully Clockwise. (Cont)
TEST OR INSPECTION
CORRECTIVE ACTION
Step 1. If continuity exists, replace component board assembly A4.
Step 2. If continuity does not exist, replace switch S3.
TABLE NO. 3-12. Multimeter Does Not Indicate 1.0 6 0.2 Vdc When CLTV STICK POSN Switch Is
Placed To SIM STAB 2.
TEST OR INSPECTION
CORRECTIVE ACTION
1. Place CLTV STICK POSN control to SIM STAB 1. Go to 2.
2. Connect multimeter between CLTV STICK POSN SIM NO. 1 and SIG GND test
jacks. Go to 3.
3. Turn CLTV STICK POSN control to a multimeter indication of 1.0 vdc. Go to 4.
4. Connect multimeter between CLTV STICK POSN SIM NO. 2 and SIG GND test
jacks. Go to 5.
5. Place CLTV STICK POSN control to SIM STAB 1&2 and then to SIM STAB 2.
Multimeter shall indicate 1.0 6 0.2 vdc when CLTV STICK POSN control is at STAB
1&2 and shall indicate 1.0 6 0.2 vdc when CLTV STICK POSN control is at SIM
STAB 2. Go to 6.
6. If multimeter indication is not correct, replace switch S3.
TABLE NO. 3-13.
SIM STAB 2 Or To SIM STAB 1.
TEST OR INSPECTION
CORRECTIVE ACTION
1. Connect multimeter between AS SENSOR SIM NO. 1 and SIG GND test jacks.
Go to 2.
2. Place AS SENSOR control to SIM STAB 2. Multimeter shall indicate 2.36 0.1 vdc.
Step 1. If multimeter indication is not correct, go to 3.
Step 2. If multimeter indication is correct, go to 4.