TM 5-6625-2691-13&P
1 - 7 . Difference in Models
auxiliary contacts used in the control circuit which
are not present on the Model A427.
This manual describes the Military Standard
Model A427 and Model 427B electrical test sets.
The Model A427 and Model A427B test sets are
e. Voltage Sensing Module. The Voltage Sensing
identical in function, capacity and operation. How-
Module on the two models is physically different,
ever, there are numerous internal changes which
and is not physically interchangeable. Model A427
makes most of the internal major assemblies non-
uses an internal control relay within the module.
Model A427B is all solid state. Also Model A427B
interchangeable. The following is a tabulated listing
of salient differences between the two models:
i n c o r p o r a t e s a time delay feature in the wind
a. Unit Enclosure. Figure 1-6 and figure 1-6.1
switch control circuit which allows the test set to
operate for about 5 seconds with the wind switch
are envelop line drawings of Model A427 and Model
open.
A427B which delineate the external package details.
f. Rectifier  Assembly.  The Model A427 test set
As illustrated the major difference is that skid run-
uses a selenium rectifier stack to generate the 28
ners were added to the bottom of Model A427B.
VDC control power required for fan operation. On
This raises the base of the unit approximately 6
Model A427B three separate silicon rectifier
inches and facilitates fork lift truck handling of the
bridges are used. Also the Model A427B requires
test set. Also, swing-down handles were added to
two 25 watt dropping resistors not used on Model
these skid runners to better facilitate hand lifting
of the test set by four men. The hand lift handles
A427. These dropping resistors are mounted with
were removed from the top of the housing as pro-
the silicon rectifiers on a single plate with a ter-
vided in Model A427. The housing assembly of the
minal strip,
two units is interchangeable.
g. Control Transformer. The Control Transform-
b. Control Panel. The control panel of Model
ers on the two models are physically different; and
A427 contains three (3) glass cartridge fuses for
not interchangeable. On Model A427B, a change in
p r o t e c t i o n of the unit control circuit. On Model
the control circuit reduced the VA capacity require-
A427B these fuses were replaced with three (3)
ment of these transformers.
thermal lag automatic circuit breakers. All other
h. Power Absorbers. The construction of the in-
controls and the method of operation of the two
dividual load elements within each Power Absorber
models are identical.
is different on the two model test sets. The electri-
c. Terminal Bracket Mounting. On Model A427
cal capacity and function of corresponding elements
the INPUT TERMINAL BOARD contains four (4)
of the two models is the same. The physical size
terminals which are .312 inch diameter studs with
and shape of the two Power Absorber assemblies is
insulating screw-on type clamping nuts for hand
also the same. However, differences in the method
tightening without tools. On Model A427B these
of electrical termination of the elements within the
Power Absorber are different. For this reason, the
studs are similar in form except they are .375 inch
P o w e r Absorbers of Model A427 are not inter-
diameter, and two studs are provided for connec-
changeable with Model A427B Power Absorbers.
tion of lead LO or neutral. The ground stud on
i. Fan Motor. The Fan Motors of Model A427
both models is identical. The input voltage select-
and Model A427B are identical, and interchange-
ing VOLTAGE RECEPTICAL is the same on both
able.
models. The operating procedures for connection
j. Wind Switch. The Wind Switch used on Model
of supply and selection of operating voltage is the
A427B is different than the switch used on Model
same for both models.
A427. The wind switches are not physically inter-
d. Contactor. T h e I N P U T P O W E R C O N T A C -
changeable.
TOR (K1) is physically different on the two
k. Variable Transformer. The Variable Trans-
models; and is not physically interchangeable. The
former of Model A427 and Model A427B are iden-
energizing coil electrical characteristics are differ-
ent in the two models. Also Model A427B has two
tical, and interchangeable.