Table 6-2. Test Set Checkout Procedure - Continued
Test instructions
Normal indication
h. Connect common lead to test point F,
positive lead to GRD. Switch S5 to
SHORT position. Set DVM to RX1OK
scale.
(1)
Switch S4 to BULK position.
(1)
50K ohms (approx )
(2)
Switch S4 to EXTERNAL
(2)
65K ohms (approx )
position.
i. Connect positive lead to test point F,
common lead to test point E. Set DVM to
RX10K scale.
(1)
Switch S4 to BULK position.
(1)
35K ohms +/-5%
(2)
Switch S4 to EXTERNAL
(2)
45K ohms +/-5%
position.
j. Connect DVM leads between test
10K ohms +/-5%
points I and J. Set DVM to RX100 scale.
k. Connect DVM leads between test
5K ohms +/-5%
points J and K. Set DVM to RX100 scale.
1.
(1)
Connect positive lead to
(1)
Infinite ohms
test point G, common lead to
GRD. Switch S5 to OPEN
position. Set DVM to RX1OK
scale. Switch S4 to BULK
position.
(2)
Connect common lead to test
(2)
Less than 100 ohms
point G, positive lead to GRD.
Set DVM to RX1 scale.
(3)
Switch S4 to EXTERNAL
(3)
Infinite ohms
position. Set DVM to RX1OK
scale.
(4)
Connect common lead to test
(4)
Less than 100 ohms
point G, positive lead to GRD.
Set DVM to RX1 scale.
6-11